dc.identifier.citation | ARAUJO, André Gustavo Rolim de. Estudo clínico observacional da facoemulsificação com ultra-som torsional na microincisão coaxial de 2,2 mm. 2008. 83 f. Dissertação (Mestrado em Ciências da Saúde)-Universidade de Brasília, 2008. | pt_BR |
dc.identifier.uri | http://repositorio.unb.br/handle/10482/1351 | - |
dc.description | Dissertação (mestrado)—Universidade de Brasília, Faculdade de Ciências da Saúde, 2008. | pt_BR |
dc.description.abstract | Objetivo: Avaliar clinicamente o desempenho da cirurgia de catarata por facoemulsificação
com tecnologia torsional Infiniti Vision SystemTM – OZilTM U/S Torsional® e incisão
microcoaxial em córnea clara de 2,2 mm.
Métodos: Estudo clínico observacional de 96 olhos com catarata nuclear ou córtico-nuclear
grau 2 de acordo com o Sistema de Classificação de Opacidades Lenticulares III, submetidos
à cirurgia de facoemulsificação com incisão microcoaxial de 2,2 mm e tecnologia torsional
OZiL (OZilTM U/S Torsional® com Infiniti Vision SystemTM no Hospital Oftalmológico de
Brasília. Foram analisadas variáveis intra-operatórias de volume de solução salina balanceada
(SSB) infundida, tempo cirúrgico (TC) e energia ultra-sônica dissipada acumulada (EDA);
variáveis pré e pós-operatórias de um dia e um mês de paquimetria central da córnea (PCC) e
densidade de células endoteliais da córnea (DCE).
Resultados: Na análise do parâmetro intra-operatório de volume médio de SSB, em ml,
utilizado durante a cirurgia, foi de 97,3 com desvio padrão (DP) de ± 9,62. O TC médio foi de
5,08 com DP de ± 0,53. A EDA média dos procedimentos foi de 3,23 com DP de ± 0,39. Os
valores médios da PCC, em μm, analisados no pré-operatório, pós-operatório com um dia e
um mês foram, respectivamente, 538,98, 545,34 e 541,78 (p-valor 0,05). Os valores médios
da DCE, em células por mm2, analisados no pré-operatório, pós-operatório com um dia e um
mês foram, respectivamente, 2321, 2281, 2230 (p valor 0,05). A redução percentual da DCE
foi de 1,7% no primeiro dia pós-operatório e 4% após um mês da cirurgia.
Conclusão: A técnica de facoemulsificação com tecnologia – OZilTM U/S Torsional® e
incisão microcoaxial em córnea clara de 2,2 mm parece não afetar prejudicialmente o
endotélio corneano, avaliado através dos parâmetros de PCC, DCE, EDA e volume de SSB
infundido.
_______________________________________________________________________________________ ABSTRACT | pt_BR |
dc.description.abstract | Objective: to clinically evaluate the performance of the phacoemulsification cataract surgery
using the torsional technology Infiniti Vision SystemTM - OZilTM U/S Torsional® with 2,2 mm
clear corneal microcoaxial incision.
Methods: observational clinical study comprising 96 eyes with nuclear or cortico-nuclear
cataract grade 2 according to Lens Opacities Classification System III submitted to
phacoemulsification cataract surgery using the torsional technology Infiniti Vision SystemTM -
OZilTM U/S Torsional® with 2,2 mm clear corneal microcoaxial incision at Hospital
Oftalmológico de Brasília. Intra-operative variables of used volume of balanced salt solution,
surgical time and accumulated dissipated ultrasound energy were analysed. Pre and one-day
and one-month post-operative variables of central corneal thickness and corneal endothelial
density were also analysed.
Results: the mean volume of used balanced salt solution, in ml, was 97,3 with standard
deviation of ± 9,62. The mean surgical time, in minutes, was 5,08 with standard deviation of ±
0,53. The mean accumulated dissipated ultrasound energy was 3,23 with standard deviation of
± 0,39. The pre-operative, one-day and one-month post-operative mean values, in μm, of
central corneal thickness were, respectively, 538,98, 545,34 and 541,78 (p value 0,05). The
pre-operative, one-day and one-month post-operative mean values of corneal endothelial
density, in cells per mm2, were, respectively, 2321, 2281, 2230 (p value 0,05). The
percentage decrease in corneal endothelial density at first post-operative day was 1,7% and at
first post-operative month was 4%.
Conclusion: the phacoemulsification technique with the torsional technology Infiniti Vision
SystemTM - OZil TM U/S Torsional® with 2,2 mm clear corneal microcoaxial incision seems
not to damage or affect negatively the corneal endothelium as suggested by the variables
analysed. | pt_BR |
dc.language.iso | Português | pt_BR |
dc.rights | Acesso Aberto | pt_BR |
dc.title | Estudo clínico observacional da facoemulsificação com ultra-som torsional na microincisão coaxial de 2,2 mm. | pt_BR |
dc.type | Dissertação | pt_BR |
dc.subject.keyword | Catarata | pt_BR |
dc.subject.keyword | Facoemulsificação | pt_BR |
dc.subject.keyword | Distúrbios da visão | pt_BR |
dc.subject.keyword | Cirurgia oftalmológica | pt_BR |
dc.location.country | BRA | pt_BR |
dc.description.unidade | Faculdade de Ciências da Saúde (FS) | pt_BR |
dc.description.ppg | Programa de Pós-Graduação em Ciências da Saúde | pt_BR |
Aparece nas coleções: | Teses, dissertações e produtos pós-doutorado
|